The facility is designed to prevent that the quarantine pathogens escape from inside. In order to achieve this, the installation has:
a) Air recirculation circuit, which is filtered through HEPA filters. When these filters become clogged, the installation must be stopped and desinfected. Then, the filters must be changed and finally the facility may be restart.
b) Three zones of negative differential pressure. When any of the doors is opened, the air comes always from outside. Thereby, the probability that the infectious agent escapes is decreased. The three zones are: entrance hall, entrance and exit corridors well differentiated and the laboratory. There is a room with a biological safety hood inside the laboratory. This room must remain closed. All the open manipulations of the infectious agent must be done inside the safety hood.
Notes: the facility is not watertight. Due to the differential pressure, there is always a flow from the outside. This also facilitates the entry of both flying and crawling insects. There is a mosquito trap and cockroach trap in the entrance hall.
- Exterior entrance door. It is always locked by the key. There is only one copy in use that must be collected and returned from the concierge.
- Entrance and exit doors to hallway and locker room. All doors are interlocked, so none of the doors can be opened if another door is already opened.
- Entrance and exit corridors. These areas act as locker rooms. In addition, they provide another differential pressure zone. The wardrobes have double doors so the objects that have been left athe the entrance, can be collected at the exit.
- Main laboratory. Contains: poyatas, shelves on poyatas, low furniture, fridge, freezer, autoclave and temperature control.
- Phytotron room. It is normally left opened to facilitate the thermal balance. It also contains a -80ºC freezer.
- Biosecurity chamber room. This room contains the chamber and a poyata. It should always be closed, especially when someone is using the chamber.
- Pile and drain. It contains a BioWaste that alerts when the tank is full.
- SAS to Exchange objects with the outside.
- Facility controls. They are external, located next to the entrance door. In a normal way, they should not be touched. Unless a maintenance is required, the facility will work continuously and the controls will not be deactivated.
1. Exterior entrance door

2-3. Entry and exit doors, hallway / locker room

4. Main laboratory

5. Phitotron room

6. Biosecurity chamber room

7. Pile and drain

8. SAS (exchang of objects with the outside)

9. Controls of the facility